Sound Waves
Sound is a pressure wave so small that a loud tone displaces air molecules by less than the width of an atom, yet a microphone reads it easily — because pressure, not displacement, is what the ear and the instrument sense. The acoustic impedance ties pressure, density, and particle velocity together, fixes the intensity a wave carries, and sets the reference for the decibel, a logarithm that tames a range in power.
╌╌╌╌
Longitudinal fields and sound speed
Sound in a gas consists of small longitudinal motions of the gas about its equilibrium state. A loudspeaker diaphragm, a tuning fork, or a vibrating surface first pushes nearby gas. That local compression raises pressure slightly. Adjacent gas responds, and the disturbance propagates through the medium. Individual molecules travel only small distances about local equilibrium positions; the sound pattern can travel many metres while the molecular excursion remains microscopic.
At a fixed instant, a longitudinal wave contains alternating regions of compression and rarefaction. Compression means that more mass occupies a small volume than in the equilibrium gas. Rarefaction means less mass occupies the corresponding volume. Both descriptions refer to departures from the ambient state. Atmospheric pressure and equilibrium density remain much larger than their oscillatory changes in ordinary sound measurements.
The same wave can be represented by several physical variables. The longitudinal displacement gives the displacement of a small material element from its equilibrium position. Particle velocity gives the time derivative of that displacement. Gauge pressure gives the pressure change above or below ambient pressure. Density change gives the corresponding density departure. The prime distinguishes a fluctuating quantity from its equilibrium value: and .
No single representation is intrinsically more complete. A microphone responds mainly to pressure. A laser vibrometer can measure surface or particle displacement. A model of energy flow often uses pressure and particle velocity together. State the measured variable before converting an amplitude or a level. A quoted “sound amplitude” without that information has no unambiguous unit or physical meaning.
A sinusoidal plane wave travelling in the positive direction has displacement form
Here is the displacement amplitude, is the wavenumber, is wavelength, is angular frequency, and is frequency. A fixed material parcel has a fixed equilibrium coordinate and oscillates back and forth. A fixed phase point, such as a pressure maximum, moves at the wave speed . Keeping those two motions separate avoids a common error: the gas does not stream from source to receiver at the sound speed.
Sound is ordinarily treated as a small-signal wave. The displacement, density change, and gauge pressure are sufficiently small that linear relations apply. Nonlinear effects become important at large pressure amplitudes, in shock waves, or very close to strongly driven sources. Unless stated otherwise, the relations below use linear acoustics.
Plane-wave language describes local behavior. A point source emits curved wavefronts, yet a small microphone diaphragm samples only a small patch of one wavefront. The patch can often be approximated as plane if its diameter is much smaller than the local radius of curvature and the wavelength-dependent spatial variation across the diaphragm is negligible.
Displacement, particle velocity, pressure, and density
Different sound variables have fixed derivative relations. Differentiate the displacement wave with respect to time:
Particle-velocity amplitude is therefore . At a displacement maximum, the particle velocity is zero. At the equilibrium crossing, the magnitude of particle velocity is largest. This is the same local phase relation as simple harmonic motion. It gives the motion of one gas parcel; wavefront propagation is a separate quantity.
Pressure follows from compression. Take two neighboring equilibrium locations, separated by . Their instantaneous separation is
The fractional length change is . In a narrow tube with fixed cross-sectional area, it also gives the fractional volume change of the gas element. Compression has : the right boundary has moved less far right than the left boundary, reducing the interval between them. Expansion has .
The bulk modulus relates a small pressure rise to fractional volume decrease:
For the sinusoidal displacement above,
The pressure amplitude is . Pressure is in phase with particle velocity for a progressive wave moving in the positive direction. Their signs are the same with the phase convention used here: a parcel moving toward positive has positive particle velocity and is part of a compression that has positive gauge pressure. Reversing propagation direction reverses the sign relation.
Mass conservation gives the density relation. A material element with a shorter length contains the same mass in less volume, so its density increases. To first order,
Combining this with the pressure relation produces . In a linear fluid, , so pressure and density departures have the same sign. The oscillatory excess density in a compression therefore coincides with positive gauge pressure.
The displacement amplitude can be tiny even when a sound is readily measured.
Microscopic displacement should not be confused with negligible transported energy; the wave transfers energy through the medium continuously.
Sound speed and the material response of a gas
The speed of a small pressure disturbance in a fluid is
where is the bulk modulus appropriate to the compression process and is equilibrium density. A large bulk modulus resists compression and raises wave speed. Greater mass density raises inertia and lowers wave speed when the modulus is held fixed. The expression has the expected dimensions because has units and has units .
The acoustic compression and expansion of a gas occur rapidly enough that little heat enters or leaves a small parcel during one cycle. The appropriate bulk modulus is therefore the adiabatic value, . The isothermal value applies to a constant-temperature compression. An ideal gas then gives
where is absolute temperature, is molar mass, is the gas constant, and . Dry air near room temperature has , , and sound speed near at .
At fixed gas composition and temperature, ideal-gas sound speed is nearly independent of ambient pressure. Both and increase proportionally with pressure, leaving their ratio unchanged. Ambient pressure still affects density, impedance, microphone loading, nonlinear behavior, attenuation, and source coupling. Sound speed and acoustic impedance therefore have different pressure dependence.
Near room temperature over a modest interval, the approximation
is a numerical check, with in degrees Celsius. The square-root expression remains the better model when temperature range, composition, or uncertainty matter. Record whether the temperature is air temperature near the propagation path or a remote weather-station value; a heated room can have a vertical temperature gradient large enough to affect a precise time-of-flight measurement.
A time-of-flight record gives estimated speed , where is sensor separation and is the propagation delay. Trigger jitter, uncertain sensor acoustic centers, reflections, and temperature gradients contribute to the uncertainty. A repeated pulse sequence permits an average delay and a scatter estimate. A sinusoidal source needs a phase method or a broadband modulation because a delay measured from a single periodic trace is ambiguous by integer periods.
Energy, intensity, and propagation loss
A sound wave transfers energy through a medium. The local energy has two interchangeable forms. Particle motion carries kinetic energy. Compression stores elastic energy because the gas has been displaced from its equilibrium volume. In a harmonic progressive wave, the two contributions vary through the cycle and have equal time averages. The mean total energy density can therefore be written in several equivalent ways:
The angle brackets denote an average over complete cycles. Subscript denotes a peak amplitude, whereas “rms” denotes root-mean-square amplitude. A sinusoid has and . Do not mix peak pressure with rms particle velocity in the same formula. A factor of two error in energy or intensity usually traces to that mismatch.
The energy density is local: it describes energy stored in one cubic metre at a particular region of the sound field. It does not by itself state how rapidly energy passes through a surface. That rate depends on wave speed and propagation direction. A plane progressive wave has time-averaged intensity
Intensity has SI units . The product is an instantaneous power flux: pressure times particle velocity gives watts per square metre. Its cycle average is positive for a progressive wave because pressure and particle velocity are in phase. In a standing-wave region, their phase relation differs, and the local cycle-averaged flux can be small even where pressure amplitude is large. The plane-progressive formulas need that condition stated whenever they are used to infer intensity from pressure.
The relation follows from a control-surface argument. During a short interval , a plane wave moves through a distance . A surface patch of area therefore receives energy from a slab of volume . If the mean energy density is , that slab contains . Dividing by produces . The argument measures energy crossing a surface normal to the direction of propagation. An oblique surface receives a smaller flux per area by the projected-area factor.
Power is intensity integrated over area. Through a surface enclosing a source,
When intensity is uniform and perpendicular to a flat receiver of area , this reduces to . A microphone pressure reading samples the local acoustic field. Total source power requires an area integration over a suitable enclosing surface.
A sinusoidal tone has intensity proportional to the square of each of these amplitudes:
Doubling pressure amplitude multiplies intensity by four. Doubling displacement amplitude at fixed frequency also multiplies intensity by four. Doubling frequency at fixed displacement amplitude multiplies particle velocity amplitude by two and intensity by four. A claim that a sound “doubled in amplitude” is incomplete unless the variable is named. A doubling of rms pressure, particle velocity, diaphragm displacement, electrical drive voltage, and emitted power are different experimental statements.
Intensity is a vector quantity before averaging over a surface. Its direction is the direction of energy transport. In a plane progressive sound wave, pressure and particle-velocity phasors align, and the mean intensity vector points along the wavevector. In a complex field, use the time average of pressure times the component of particle velocity normal to the surface. A pressure-only microphone cannot determine intensity direction by itself; a paired pressure–particle-velocity probe or an array measurement resolves the missing directional information.
The average power of a source does not equal the electrical power printed on a loudspeaker amplifier. Electrical input power divides among acoustic radiation, heating in the voice coil, mechanical loss, enclosure vibration, and circuitry. Acoustic output power must be inferred from acoustic measurements on a specified surface or reported from a calibrated source characterization. Name which power is being reported: electrical input, acoustic radiation, or power incident on a receiver.
Geometric spreading, point sources, and propagation loss
A compact source in an open region can be approximated by a point source if the receiver distance is large compared with the source dimensions. If it radiates uniformly in every direction and loses negligible energy to absorption, the same acoustic power crosses every sphere centered on the source:
The inverse-square relation comes from geometric spreading. A sphere with twice the radius has four times the area, so the same power is distributed over four times the surface. It does not describe molecular energy loss from the wave; it describes an area increase. Absorption adds an additional decrease whose magnitude depends on frequency, humidity, gas composition, and path length.
The level change between two distances follows immediately:
Doubling distance from an ideal point source reduces intensity by a factor of four and reduces intensity level by approximately . Ten times the distance reduces level by . These values apply to a single free-field source. They cannot be transferred unchanged into a room where direct sound combines with reflected sound.
A source mounted in a large rigid plane radiates primarily into a hemisphere. With uniform hemispherical radiation,
At a given distance and total acoustic power, the hemispherical intensity is twice the isotropic full-sphere value. A loudspeaker, a mouth, or a machine enclosure usually has its own directional pattern. A report should state the direction and distance of every microphone relative to the source axis.
Absorption and obstacles alter the point-source model. A phenomenological attenuation form is
when is an attenuation coefficient expressed in decibels per metre for a specified frequency and propagation condition. This representation separates geometric spreading from path attenuation. It should be fitted only over a region where source directionality and reflection conditions are controlled. An arbitrary level-versus-distance slope does not establish an absorption coefficient.
Near a source, the measured field can contain reactive energy that alternately stores and returns energy to the source. The pressure-to-velocity ratio then differs from , and pressure alone does not give a reliable intensity estimate. The transition distance depends on source size and wavelength. Avoid claiming free-field intensity from a microphone placed against a loudspeaker grille, inside a duct termination, or a few centimetres from a vibrating panel without an appropriate near-field model.
An open-space measurement also needs a background record. Turn the source off while leaving microphone gain, geometry, and time weighting unchanged. If the background level is close to the source-on level, subtraction must be done with linear mean square pressure or intensity, never by arithmetic subtraction of decibel values. The later decibel section gives the needed conversion.
Decibel levels and acoustic measurements
Sound intensity spans an enormous physical range. A logarithmic ratio compresses that range while preserving multiplicative changes. The intensity level is defined relative to a reference intensity by
In air-acoustics examples, is commonly used as the reference intensity. A sound at has level on that scale because its intensity ratio is . The decibel is a ratio label. It does not carry the dimensions of watts per square metre, pascals, or acoustic power.
The factor of 10 in the definition follows the use of an energy-like quantity. For two intensity values,
An intensity ratio of 2 gives , a ratio of 10 gives , and a ratio of 100 gives . A change means one tenth of the original intensity. The sign describes a ratio; it does not mean that physical intensity became negative.
Microphones measure pressure, so sound-pressure level is often more direct:
The conventional reference pressure in air is . The factor is 20 instead of 10 because intensity is proportional to pressure squared in a progressive plane wave. When both references are matched by , sound-pressure level and intensity level have nearly equal numerical values under plane-wave free-field conditions.
Pressure level does not universally equal intensity level. The relation can fail near boundaries, in small rooms, in ducts, and close to sources because acoustic impedance is not necessarily the real quantity . A pressure meter then still reports a valid pressure level, but it does not alone establish net intensity or acoustic power. Record the quantity measured instead of relabeling every microphone reading as intensity.
The phrase “twice as loud” belongs to perception and cannot be read directly from one decibel difference. Loudness depends on frequency, duration, spectral content, listener, and listening environment. Here decibels describe a physical pressure or intensity ratio. A report can separately state an A-weighted sound level or a subjective listening result, provided the physical quantity and weighting are identified.
Combining independent sources
Independent sources add in linear intensity or mean-square pressure. Their decibel levels must be converted before addition:
For equal, independent sources at the same receiver location,
Two equal sources give a increase. Ten equal sources give a increase. Two sources with levels differing by more than about produce a total only slightly above the stronger source because the lower linear intensity is a small fraction of the total.
Coherent sources require a pressure-phase calculation. For two harmonic tones at the same frequency, the instantaneous pressures add:
Equal pressures with the same phase produce twice the pressure amplitude and four times the intensity, a increase. Equal pressures in opposite phase cancel in the ideal model. These cases depend on stable relative phase at the receiver. Independent noise sources, separate machines with drifting phase, and long-time energy averages use intensity addition.
Background correction is another addition problem. Let be the level with source and background present, and let be the level with the target source absent. If the sources are independent over the measurement average,
The subtraction occurs inside the linear sum. If source-on level is only slightly above background, the difference of two nearly equal quantities has large relative uncertainty. Report the source-on level and background level along with the corrected result. A correction derived from a short quiet interval cannot represent an intermittent background such as traffic, ventilation cycling, or speech.
Free-field point sources permit direct geometric-spreading and level arithmetic.
The prediction is a model result, not a calibration certificate. It needs a source axis, an open propagation path, comparable environmental conditions, and a far-field distance range. A measurement that differs from the prediction may indicate directionality, a reflecting surface, air attenuation, source-output drift, or microphone placement error. Retain the residuals alongside the ideal curve.
Microphone measurements and calibration
A microphone converts an acoustic pressure variation at its diaphragm into an electrical signal. Its sensitivity is commonly specified in volts per pascal at a reference frequency and incidence condition. If a calibrated microphone has sensitivity and the electronics have voltage gain , a sinusoidal pressure with rms value produces an rms output approximately
The approximation omits several instrument properties: microphone frequency response, phase response, directional response, preamplifier gain, output impedance, noise, and maximum input pressure. A specification sheet does not replace calibration under the measurement configuration. The signal path must be known from diaphragm to stored sample.
Broadband-measurement sensitivity depends on frequency. Let be the complex frequency response including amplitude and phase. A sampled voltage spectrum corresponds to pressure spectrum only within the calibrated bandwidth. Dividing by a response value outside its valid range can strongly amplify electrical noise. A measurement report should give the frequency band retained in the analysis and state whether a response correction was applied.
The sensor itself can be characterized for several acoustic conditions. A pressure response applies when the diaphragm is flush with a boundary or inside a suitable coupler. A free-field response applies for a specified incidence direction in an approximately reflection-free wave. A random-incidence response averages over many arrival directions. These calibration conditions have different corrections at high frequency because diaphragm geometry and acoustic scattering matter. Copying a free-field calibration into a boundary measurement changes the pressure estimate.
Establishing the pressure scale
A field calibration uses a source that produces a known rms pressure at the microphone. If a calibrator has pressure level at a stated frequency, the known pressure is
With observed calibrator voltage , the system scale factor is
This factor includes microphone sensitivity and the active gain setting. It is valid only while the same gain, input range, coupling, and meter mode remain in use. A pre-run and post-run calibration comparison can identify drift, loose connectors, battery loss, or an accidental gain change. If the two values disagree beyond the instrument tolerance, retain both records and determine whether the run requires repetition or a stated calibration uncertainty.
The numerical neatness arises from the selected values. In a real measurement, keep more digits internally and round the reported level only after evaluating the uncertainty. The calibration tone frequency is also part of the record because a single-frequency calibration cannot establish an entire frequency response.
An rms meter uses a finite averaging interval. The rms pressure of a time record over interval is
The averaging interval must contain enough cycles to stabilize a tonal rms estimate. At , a window contains only two cycles; phase placement can noticeably change an unwindowed estimate. At , the same interval contains one hundred cycles. Broadband and impulsive sounds need a window length chosen for the physical question, then stated with the result.
Clipping cannot be repaired by applying a calibration factor after the fact. The flattened waveform has lost amplitude and spectral information. Set the input range with a short test recording or use an instrument with documented headroom. Electrical noise also needs attention. A microphone self-noise specification and preamplifier noise combine with acoustic background. Measure the quiet-system level with the microphone installed and the source absent; determine whether the desired signal exceeds that floor by a sufficient margin for the claimed uncertainty.
Geometry, orientation, and spatial sampling
Acoustic results are functions of position. A microphone location requires three coordinates, source distance, source-axis angle, height above the floor, and the positions of nearby large boundaries. “One metre from the speaker” is incomplete if the microphone could have been on-axis, behind the source, near a desk, or beside a wall. A tape measure gives distance; it does not determine the acoustic geometry.
Spatial sampling interval depends on wavelength and the purpose of the scan. A single-source far-field level may use one carefully documented point. Mapping a reflected sound field requires multiple locations because pressure extrema can occur over distances comparable with a fraction of a wavelength. Higher frequencies have shorter wavelengths and require finer spatial spacing. The scan path should be marked physically or located with coordinates; a hand-held microphone moved informally through space cannot support a reproducible spatial map.
Microphone orientation matters when sensitivity varies with incidence angle. Align a free-field microphone according to its calibration convention. Keep the operator and support hardware out of the direct path when possible, because body reflection and scattering can change local pressure. A tripod reduces motion and makes post-run repositioning possible. Record whether a windscreen, extension cable, boundary mount, or protective grille was installed because each can alter high-frequency response.
Repeated measurements separate source variation from meter scatter. At each planned location, record several independent time windows, then report mean level, temporal range, and the averaging rule. If the source output changes over time, randomize or cycle measurement positions so that a position trend is not confused with source drift. A measurement program with one reading per position cannot distinguish those two causes.
Exposure, data records, and quality checks
Intensity describes instantaneous average power flow per area. Acoustic exposure adds that flow over time:
Constant progressive-wave intensity gives . Doubling duration doubles exposure even though the instantaneous level is unchanged. A high-level short event and a lower-level long event can therefore have comparable integrated energy. Duration belongs in any exposure statement.
Equivalent continuous sound-pressure level compresses a varying pressure record into one level with the same mean-square pressure over a specified interval:
The interval is part of the quantity. A one-minute equivalent level and an eight-hour equivalent level are different summaries even if their numerical values happen to match. A time-weighted maximum, a percentile level, a peak pressure, and an equivalent level answer different questions. Do not replace one with another without saying so.
Frequency weighting is a separate choice. An unweighted record, often called Z-weighted, retains the instrument response over its stated band. An A-weighted record applies a frequency-dependent weighting before level integration. The suffix must be preserved in notation: and do not mean the same physical processing. A result reported only as “dB” leaves the weighting, time behavior, and reference quantity unspecified.
Measurements of exposure should retain raw or sufficiently detailed time data when possible. A sole daily equivalent level cannot distinguish continuous machinery noise from a small number of impulses or a sequence of tonal events. A compact report includes at least a time history or time-binned level table, the equivalent level, the maximum detector mode, total interval, and any frequency weighting. Those records allow later comparison without reconstructing the experiment from memory.
Required record fields
- Source state: source identity, drive condition, operating cycle, and any output monitor.
- Environment: air temperature, relative humidity when relevant, propagation space, nearby boundaries, and background-source state.
- Geometry: source coordinates and axis, microphone coordinates and orientation, surveyed distance, mounting method, and scan order.
- Instrument chain: microphone model or sensitivity file, calibration method, calibrator value, preamplifier gain, input range, sample rate, and analysis band.
- Level definition: pressure or intensity quantity, reference value, frequency weighting, time weighting, rms or peak convention, and averaging interval.
- Uncertainty: repeatability, calibration tolerance, background contribution, position tolerance, and any model assumptions used for power or range conversion.
An uncertainty statement should follow the calculation path. For the calibrated voltage example, let the calibration factor have relative standard uncertainty , and let the rms voltage have relative standard uncertainty . If they are independent, the pressure relative uncertainty is approximately
For small relative changes, the corresponding pressure-level uncertainty is
A 5% relative pressure uncertainty gives approximately . That conversion is local; larger uncertainties should be propagated through the logarithm directly. A point-source power estimate also carries uncertainty from distance, directionality, range selection, and environmental variation. Meter precision alone does not quantify model uncertainty.
The measurement objective determines the required detail. A classroom demonstration may need calibrated pressure at one documented point. A source-power estimate needs an enclosing-surface design or a justified radiation model. A room survey needs a spatial grid and a reflection-aware interpretation. An exposure record needs duration, weighting, time integration, and source-state history. The same microphone can serve each task, but the geometry, processing, and uncertainty model differ.
Pressure survey and quality checks
A loudspeaker operates with a stable electrical drive in an open indoor space. The measurement goal is a pressure-level map along one documented source axis, followed by an exposure summary for a ten-minute operating interval. The goal does not include a claim of total acoustic power, because a single axis scan does not measure radiation over an enclosing surface. Stating that boundary at the outset prevents a pressure survey from being reported as a power characterization.
Mount a calibrated microphone on a tripod with its reference point at the same height as the loudspeaker acoustic center. Survey positions at , , and along the forward axis. Measure the temperature near the path, retain the source-drive monitor, and acquire a source-off background interval at every position. Revisit the position after each new range. That return measurement tracks output drift while leaving the range sequence explicit.
Suppose the calibrated one-minute source-on levels are , , and at the three ranges. The corresponding background levels are , , and . Background matters increasingly at larger range because the target source becomes less dominant. Convert each pair to linear ratios before subtraction:
At , the correction from background-inclusive level is small because background is lower. At , the difference means background occupies a large part of the source-on mean-square pressure. The corrected source estimate has much larger uncertainty at that range even if the meter display has the same resolution.
The ideal free-field point-source trend from to is ; from to it is . Compare those predictions with background-corrected levels, not with uncorrected values. If the corrected results depart from the ideal trend, list likely experimental causes in a separate interpretation field: wall reflection, source directivity, source heating, geometry error, background variability, or insufficient far-field range. Do not insert an absorption coefficient from the three-point slope difference alone; the fitted slope must be compared with its uncertainty and the stated propagation model.
The calibration check can be processed on the same scale. Assume the pre-run calibrator check reads after applying the system factor and the post-run check reads . The difference is a traceable drift indicator. Take half the difference, or a documented alternative based on laboratory practice, as one contribution to calibration uncertainty. It should not be silently adjusted away by forcing both checks to read the nominal value.
For the ten-minute exposure summary, acquire a continuous pressure time record with the selected weighting documented. Calculate from the full interval. Segment the record into one-minute values as a diagnostic, then compare their linear mean-square average with the ten-minute result. A large difference between segments signals a changing source or background. The single ten-minute equivalent level remains a valid energy-equivalent summary, while the one-minute sequence preserves temporal variation needed for interpretation.
If the measurement uses A weighting, apply it before forming the mean-square integral and report . If the goal is a physical pressure comparison with a plane-wave intensity conversion, retain an unweighted band-limited pressure record as well. A-weighted values are valuable summaries of weighted content, but they are not substitutes for raw pressure when an impedance or power calculation is required.
Before publication or comparison, perform four final checks. Verify that the source state did not change across the relevant time interval. Verify that pre-run and post-run calibration values agree within the stated uncertainty. Verify that no stored waveform is clipped and that the source signal remains adequately above background and instrument noise. Verify that each stated conversion—pressure to intensity, range extrapolation, or acoustic power—has the geometry and propagation conditions needed by its model. These checks connect a displayed decibel value to a measurement that another reader can assess, reproduce, or challenge quantitatively.
Measurement model boundaries
The common acoustic relations apply under different physical conditions. Keeping those conditions attached to the formula is part of accurate reporting. The relation describes a small disturbance in a material with a specified bulk response. The ideal-gas form uses an adiabatic compression and a uniform gas composition. The plane-wave relation requires a locally progressive wave and a pressure–particle-velocity relation close to the real impedance . The point-source relation needs a far-field range, known radiation geometry, and a path where reflections and absorption are either negligible or explicitly modeled.
The limitations do not invalidate a measurement. They identify the quantity that was actually measured and the additional data required for a conversion. A pressure reading beside a loudspeaker can compare repeatable drive settings. It becomes an intensity estimate only after pressure–velocity behavior is established. A range scan in a room still documents the received pressure distribution. It becomes an inverse-square test only after direct and reflected contributions are separated or controlled. This distinction preserves the experimental result while preventing unsupported claims.
Temperature enters several stages of the analysis. It changes sound speed through the gas equation, affects wavelength at a fixed source frequency, and can alter air density in the impedance used for a pressure-to-intensity conversion. For small temperature changes in an ideal gas,
A careful report records the temperature and the temperature location. It need not apply a correction when the intended result is measured sound-pressure level at that condition. It does need a correction or uncertainty allowance when comparing time-of-flight speed, wavelength, or intensity estimates across materially different conditions.
The same discipline applies to decibels. A pressure level is a logarithmic pressure ratio. An equivalent level is a time-integrated mean-square-pressure statistic. An intensity level is a logarithmic power-flux ratio. An A-weighted level includes a frequency weighting. Each remains meaningful when its reference, interval, and processing are named. Ambiguity enters when a bare number is moved between those definitions without the conditions that make the conversion valid.
╌╌ END ╌╌